Model |
KS-210 Spectrophotometer |
|
Illumination/Viewing Geometry |
D/8 (Diffuse illumination, 8° viewing), SCI/SCE (Specular Component Included/Excluded). Compliant with CIE No.15, GB/T 3978, GB 2893, GB/T 18833, ISO7724-1, ASTM E1164, DIN5033 Teil7. |
|
Features |
Utilizes a dual-path silicon photodiode array. Ideal for color difference quality control in industries such as plastics, electronics, paints and inks, textile dyeing and printing, printing, and ceramics. |
|
Light Source |
Combined full-spectrum LED light source. |
|
Integrating Sphere Size |
Φ40mm |
|
Spectral Device |
Nano-integrated spectral device |
|
Sensor |
Silicon photodiode array (dual-row, 18 groups) |
|
Wavelength Range |
400-700nm |
|
Wavelength Interval |
10nm |
|
Measurement Aperture |
6 Apertures:3mm platform+3mm pointed tip+5mm platform+5mm pointed tip+10mm platform+10mm pointed tip |
|
Specular Mode |
Simultaneous measurement SCI/SCE |
|
Color Space |
CIE LAB,XYZ,Yxy,LCh,CIE LUV,s-RGB,HunterLab,βxy,DIN Lab99 |
|
Color Difference Formula |
ΔE*ab,ΔE*uv,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00,DINΔE99,ΔE(Hunter) |
|
Other Color Indices |
Spectral Reflectance, WI (ASTM E313, CIE/ISO, AATCC, Hunter, Taube, Berger, Stensby), YI (ASTM D1925, ASTM 313), Metamerism Index MI, Staining Fastness, Shade Change Fastness, Strength, Opacity, 555 Shade Classification, Munsell (C/2), Blackness (My, dM), Color Density CMYK, Tint (Some functions realized via PC software) |
|
Observer Angle |
2°/10° |
|
Calibrated Illuminants |
D65,A,B,C,D50,D55,D75,F1 ,F2(CWF),F3,F4,F5,F6, F7(DLF),F8,F9,F10(TPL5),F11(TL84), F12(TL83/U30),U35,NBF,ID50, ID65,LED-B1,LED-B2,LED-B3,LED-B4, LED-B5,LED-BH1,LED-RGB1, LED-V1,LED-V2(Partially realized via PC software) |
|
Display Data |
Spectral graph/data, sample color values, color difference values/graph, pass/fail results, color simulation, color bias |
|
Measurement Time |
Approx. 1 second |
|
Repeatability |
Chromaticity Value: Within ΔE*ab 0.028 (Standard deviation) for MAV/SCI (after warm-up and calibration, average of 30 readings of white board at 5s intervals).
Spectral Reflectance: Within 0.1% standard deviation for MAV/SCI (within 0.2% for 400-700nm). |
|
Inter-Instrument Error |
Within ΔE*ab 0.35 for MAV/SCI (average of 12 BCRA Series II color tiles). |
|
Displayed Accuracy |
0.01 |
|
Reflectance Range |
0-200% |
|
Reflectance Resolution |
0.01% |
|
Measurement Mode |
Single measurement, average measurement (2-99 times). |
|
Targeting |
Stabilizer plate positioning + Camera framing. |
|
White Calibration |
Contact-type white calibration. |
|
Dimensions |
120*75*207mm |
|
Weight |
367g(without calibration base) |
|
Battery |
Lithium-ion battery, 3.7V, 3200mAh, supports 8000 measurements within 8 hours. |
|
Light Source Life |
Over 1.2 million measurements for 10 years. |
|
Display |
TFT true color 3.5-inch capacitive touch screen. |
|
Interface |
USB,Bluetooth |
|
Data Storage |
500 standards, 10,000 samples (SCI/SCE data included in one record). Mass storage via APP/PC. |
|
Software Support |
Android, iOS, Windows, WeChat Mini-Program, HarmonyOS. |
|
Languages |
Simplified Chinese, English, Traditional Chinese |
|
Operating Environment |
Temperature: 0~40℃, Humidity: 0~85% RH (non-condensing), Altitude: below 2000m. |
|
Storage Environment |
Temperature: -20~50℃, Humidity: 0~85% RH (non-condensing). |
|
Accuracy Guarantee |
Guaranteed metrological certification. |
|
Standard Accessories |
Power adapter, data cable, user manual, quality management software (download from official website), white calibration box, protective cover, wrist strap, measurement apertures. |
|
Optional Accessories |
Micro-printer, powder test box. |
|
Note |
Specifications are subject to change without prior notice. |